Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Thermal stability testing of chemically active substances is a critical safety evaluation used to identify the temperatures at which reactive chemicals begin to decompose, release heat, or generate hazardous pressure. By applying standardized techniques such as differential scanning calorimetry, thermogravimetric analysis, adiabatic calorimetry and heat storage tests, manufacturers and shippers can determine key safety parameters including decomposition onset temperature and the self-accelerating decomposition temperature (SADT). This article explains the principles, sample types, measured parameters, applicable standards and practical applications of thermal stability determination for reactive chemicals.

Technical Overview

Chemically active substances, including self-reactive chemicals, organic peroxides, energetic materials and other thermally unstable compounds, have the potential to undergo exothermic decomposition, self-accelerating reaction or runaway reaction when exposed to heat. Thermal stability determination is the systematic measurement of how a substance behaves under controlled thermal stress, providing quantitative data on the onset of decomposition, the amount of heat and gas released, and the kinetics of the reaction.

The core objective of thermal stability testing is to establish safe temperature limits for production, storage, handling and transport. If the temperature of a reactive substance rises above a critical threshold, heat generated by decomposition may exceed the heat removed by the surroundings, resulting in self-acceleration, pressure buildup, fire or explosion. Testing is therefore an essential element of process safety studies, dangerous goods classification under the UN Model Regulations and the Globally Harmonized System (GHS), and compliance with transport and storage requirements.

Thermal stability assessment is typically performed in a tiered manner, beginning with small-scale screening tests using milligram samples, followed by larger-scale adiabatic or heat storage tests on gram to kilogram quantities when screening indicates significant exothermic potential. Internationally recognized methods are published by ISO, ASTM, EN, GB, JIS and the United Nations Committee of Experts on the Transport of Dangerous Goods.

Testing Samples

Self-Reactive Substances

Solid or liquid substances that decompose exothermically without involvement of oxygen, including azo compounds, diazonium salts, hydrazine derivatives and nitroso compounds. Thermal stability data directly determine their classification and packaging conditions.

Organic Peroxides

Peroxoxides used as polymerization initiators, curing agents and bleaching agents are among the most thermally sensitive commercial chemicals. Their SADT values determine whether temperature-controlled transport is mandatory.

Energetic Materials

Explosives, propellants, pyrotechnic compositions and their intermediates require demonstrated stability under prolonged heat exposure before transport and storage are permitted.

Pharmaceuticals and Fine Chemicals

Active pharmaceutical ingredients, intermediates, dyes and catalysts are screened for latent exotherms that may be triggered during drying, distillation, milling or long-term storage.

Polymerizable Monomers and Oxidizers

Monomers such as acrylates, styrene and vinyl compounds may undergo uncontrolled polymerization; oxidizing substances and oxidizing mixtures are evaluated for thermal compatibility and safe storage temperatures.

Testing Items

Thermal stability determination typically reports the following parameters:

  • Onset temperature of decomposition (extrapolated onset of the first exothermic event)
  • Peak decomposition temperature and peak heat flow
  • Decomposition enthalpy (heat of reaction, expressed in J/g)
  • Mass loss behavior as a function of temperature
  • Self-accelerating decomposition temperature (SADT)
  • Adiabatic temperature rise and time to maximum rate of decomposition (TMR, including TMR24)
  • Kinetic parameters such as activation energy and pre-exponential factor
  • Gas generation and pressure development rates under closed conditions
  • Constant-temperature stability and mass loss during long-duration heat storage

Testing Methods

Differential Scanning Calorimetry (DSC) and Differential Thermal Analysis (DTA)

DSC and DTA are the primary screening techniques. Milligram-scale samples are heated at controlled rates, commonly 2 to 10 K/min, in open or sealed high-pressure crucibles under an inert atmosphere. The methods detect exothermic and endothermic events and quantify decomposition energy. Relevant standards include ASTM E537, GB/T 22232-2008, GB/T 19466, ISO 11357 and JIS K 0129.

Thermogravimetric Analysis (TGA)

TGA measures mass change as a function of temperature or time and is used alongside DSC to distinguish decomposition, desolvation, drying and sublimation. ISO 11358 provides the general framework for thermogravimetric measurements.

Adiabatic Calorimetry

Adiabatic calorimeters use a heat-wait-search protocol to detect the onset of self-heating at temperatures well below those observed in DSC, then track self-heating rate and pressure development under near-adiabatic conditions. After correction for the thermal inertia of the test cell, the data yield the adiabatic temperature rise, time to maximum rate and temperature at which TMR equals 24 hours, which are key inputs for process safety evaluation. ASTM E1981 provides guidance for this technique.

Kinetic Analysis

Multiple heating rate experiments combined with isoconversional, model-free kinetic methods allow extrapolation of decomposition behavior to lower temperatures and longer times. ASTM E698 describes the determination of Arrhenius kinetic constants from DSC data, and kinetic parameters are frequently used to calculate SADT for defined package sizes.

Constant-Temperature and Heat Storage Tests

Isothermal exposure tests, Dewar vessel heat storage tests and oven storage tests evaluate whether a packaged substance remains stable over extended periods, typically several days, at a specified temperature. ASTM E487 covers constant-temperature stability of chemicals, GB/T 13464-1992 describes thermal analysis methods for material thermal stability, and EN 13631-2 specifies the determination of thermal stability for civil-use explosives. Acceptance is based on criteria such as absence of self-heating, ignition or explosion and limited mass loss.

SADT Determination (UN Test Series H)

The UN Manual of Tests and Criteria defines test series H for determining the self-accelerating decomposition temperature of substances in their actual commercial packaging. The selected test method depends on the physical state of the substance and the package size, and the resulting SADT value governs the assignment of control and emergency temperatures for temperature-sensitive goods.

Applications

  • Dangerous goods classification and packaging assignment for self-reactive substances and organic peroxides under UN, GHS and national transport regulations
  • Derivation of control and emergency temperatures for temperature-controlled storage and transport
  • Thermal runaway hazard assessment and definition of safe process temperatures in chemical manufacturing
  • Design of emergency pressure relief systems based on adiabatic pressure and gas generation data
  • Warehouse safety management, including segregation, ventilation and maximum storage temperature limits
  • Shelf-life studies, stability assessment for registration dossiers and quality control of reactive intermediates
  • Evaluation of thermal compatibility of mixtures, contaminated or reclaimed materials before reuse or disposal

FAQ

What is the self-accelerating decomposition temperature (SADT)?

SADT is the lowest ambient temperature at which self-accelerating decomposition of a substance in its commercial packaging occurs within a defined period, normally seven days. Substances with a low SADT must be transported and stored under temperature control with assigned control and emergency temperatures.

How much sample is required for testing?

DSC and TGA screening requires only a few milligrams per run. Adiabatic calorimetry typically uses grams of material, while heat storage and SADT tests may require hundreds of grams to several kilograms depending on package configuration and test design.

What is the difference between DSC and adiabatic calorimetry?

DSC uses tiny samples under programmed heating to quickly screen for exotherms and quantify energy release, but its sensitivity is limited by the scanning rate. Adiabatic calorimetry uses larger samples held under near-adiabatic conditions, allowing detection of much weaker self-heating and providing realistic pressure and runaway behavior for safety design.

How are results used to set safe temperature limits?

A common engineering practice is to keep operating and storage temperatures well below the detected decomposition onset, with margins verified by time-to-maximum-rate analysis. For packaged reactive goods, control temperatures are set below the SADT according to the rules of the UN Model Regulations.

Which standards govern thermal stability testing?

Frequently applied standards and references include ASTM E537, ASTM E487, ASTM E698, ASTM E1981, ISO 11357, ISO 11358, GB/T 22232-2008, GB/T 13464-1992, GB/T 19466, JIS K 0129, EN 13631-2 and the UN Manual of Tests and Criteria, test series H. Accredited third-party laboratories select the appropriate method based on the substance state, intended use and regulatory context.