Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Oxygen permeability testing is a critical quality control procedure for evaluating the gas barrier performance of plastic films and sheets used in food, pharmaceutical, and industrial packaging. This technical article explains the principles, standard test methods (ASTM D3985, ISO 15105, GB/T 19789), sample requirements, and typical applications of oxygen transmission rate (OTR) determination for plastic sheeting materials.

Technical Overview

Oxygen permeability determination measures the rate at which oxygen molecules migrate through plastic films and sheeting under controlled conditions of temperature, humidity, and partial pressure difference. This property, commonly expressed as the Oxygen Transmission Rate (OTR), is a fundamental barrier characteristic that directly determines the shelf life of oxygen-sensitive products, the protective performance of packaging structures, and the suitability of materials for modified atmosphere packaging (MAP) and other advanced packaging technologies.

Plastic sheeting materials exhibit oxygen transmission rates spanning several orders of magnitude. Non-polar polyolefins such as low-density polyethylene (LDPE) typically show OTR values in the range of 2,000–8,000 cm³/(m²·24h), while high-barrier materials such as ethylene-vinyl alcohol copolymer (EVOH) and polyvinylidene chloride (PVDC) coatings can achieve values below 1 cm³/(m²·24h). Aluminum foil laminates and metallized films can reach even lower values approaching virtual oxygen impermeability. Quantifying these differences with precision requires standardized instrumental methods performed under strictly controlled environmental conditions.

The most widely referenced standards for this determination include ASTM D3985 (coulometric sensor method), ISO 15105-1 (differential-pressure method), ISO 15105-2 (equal-pressure method), GB/T 19789 (Chinese national standard, coulometric method), and JIS K 7126. Selection of the appropriate method depends on the material type, expected transmission range, and the influence of humidity on the polymer matrix.

Testing Samples

Monolayer Plastic Films and Sheets

Single-layer materials such as LDPE, LLDPE, HDPE, PP (polypropylene), PET (polyethylene terephthalate), PVC, PA (polyamide/nylon), PS, and PLA sheeting represent the most common sample category. Samples should be flat, free of creases, pinholes, inclusions, and thickness variations, with a minimum testing area typically defined by the instrument's diffusion cell geometry (commonly 5 cm² to 50 cm²). A minimum of three specimens per material is recommended for statistical validity.

Multilayer and Laminated Structures

Coextruded films, adhesive laminates, and coated structures (for example, PET/EVOH/PE, OPP/PVDC, BOPP/metallized layers) are tested to verify the barrier contribution of each functional layer in the composite structure. Specimen preparation must avoid delamination at the edges, and cut edges should be cleanly sealed during mounting to prevent lateral oxygen ingress between layers.

Coated and Treated Materials

Silicon oxide (SiOx) and aluminum oxide (AlOx) coated films, metallized films, and films with organic barrier coatings require careful handling to prevent abrasion or cracking of the thin inorganic coating. The coated side orientation should be documented, as asymmetric structures may exhibit direction-dependent transmission behavior.

Sample Conditioning Requirements

Prior to testing, specimens must be conditioned in a controlled environment, typically 23°C ± 2°C and 50% ± 5% relative humidity for a minimum of 48 hours, unless otherwise specified by the relevant standard or material specification. Conditioned samples should be handled with gloves to prevent contamination of the test area by skin oils.

Testing Items

The core measurable parameters in oxygen permeability determination include:

  • Oxygen Transmission Rate (OTR): the volume of oxygen passing through a unit area of the specimen per unit time, expressed in cm³/(m²·24h) under specified test conditions.
  • Oxygen Permeability Coefficient (P): a material-intrinsic property normalized for thickness, expressed in cm³·cm/(m²·24h·cmHg) or cm³·mm/(m²·24h·atm), used to compare materials independently of gauge.
  • Oxygen Permeance: the OTR normalized by the oxygen partial pressure difference across the specimen, expressed in cm³/(m²·24h·atm).
  • Temperature dependence characteristics: OTR values at multiple temperatures for activation energy analysis using Arrhenius relationships.
  • Humidity dependence characteristics: OTR values at controlled relative humidity levels (typically 0%, 50%, 85%, and 100% RH) to characterize moisture-sensitive barrier polymers such as EVOH and polyamide.

Testing Methods

Coulometric Sensor Method (Equal-Pressure Method) — ASTM D3985 / ISO 15105-2 / GB/T 19789

In the coulometric method, the specimen is mounted as a sealed barrier between two chambers. One side is swept with pure oxygen (test gas, ≥99.5% purity), while the opposite side is swept by oxygen-free carrier gas (typically a nitrogen-hydrogen mixture). Oxygen that permeates through the specimen is entrained by the carrier gas and delivered to a coulometric sensor, where it is electrochemically reduced. The electrical current generated is directly proportional to the mass of oxygen reaching the sensor, following Faraday's law. Because both sides of the specimen are at essentially equal total pressure, this method eliminates the risk of leakage-driven errors and pressure-induced deformation, making it the globally preferred method for flexible packaging materials.

Differential-Pressure Method (Manometric Method) — ISO 15105-1 / ASTM D1434 / JIS K 7126

In the pressure-difference method, one side of the specimen is exposed to oxygen at a known pressure while the other side is evacuated to a low residual pressure. Oxygen permeating the specimen increases the pressure on the low-pressure side, and this pressure rise is measured over time with high-sensitivity pressure transducers. The transmission rate is calculated from the steady-state rate of pressure change. This method requires no carrier gas and no sensor chemicals, but it imposes a total pressure differential across the sample and demands rigorous leak checking, vacuum integrity, and temperature stability.

Standard Test Conditions

Tests are typically conducted at 23°C ± 1°C (or 23°C ± 2°C depending on the standard) with oxygen at approximately 1 atm test pressure on the high side. Relative humidity conditions may range from 0% RH (dry) to 90% RH or higher for humidity-conditioned evaluations. The instrument chamber temperature must be controlled within ±0.1°C to ±0.5°C, as oxygen permeability of polymers approximately doubles for every 7–10°C temperature increase.

Test Procedure Summary

  • Measure and record specimen thickness at multiple points; verify uniformity within ±10%.
  • Condition specimens according to the specified atmosphere.
  • Mount the specimen securely in the diffusion cell with a hermetic seal (typically using vacuum grease or a gasket system); verify the absence of leaks.
  • Purge both chambers with conditioning gases until a stable baseline is established.
  • Introduce oxygen to the test side and monitor the sensor signal until steady-state transmission is reached.
  • Calibrate the instrument periodically using a reference material (such as NIST-traceable polyester film or certified reference film) with a known OTR value.
  • Calculate OTR, permeance, and permeability from the steady-state signal; report all test conditions, specimen thickness statistics, and mean values with standard deviations.

Applications

Oxygen permeability data for plastic sheeting support a wide range of industrial and regulatory applications:

  • Food packaging development: selecting barrier structures for oxygen-sensitive foods such as processed meats, dairy products, nuts, coffee, dried foods, and beverages; validating MAP and vacuum packaging performance.
  • Pharmaceutical packaging: qualifying blister films, sachets, and laminate structures for tablets, effervescent products, and oxidation-sensitive formulations, in alignment with stability study requirements.
  • Medical device packaging: verifying sterile barrier performance of breathable and non-breathable packaging materials per recognized sterilization compatibility frameworks.
  • Material R&D and comparison: evaluating EVOH grades, nanocomposite barrier materials, and coated films; ranking candidate structures during development.
  • Quality control and supplier verification: incoming material inspection, lot-to-lot consistency monitoring, and process change validation in manufacturing.
  • Shelf-life modeling: incorporating OTR values into predictive models that estimate product oxidation over the intended storage period.
  • Compliance and certification: accredited third-party laboratories provide test reports accepted by retailers, brand owners, and regulatory submissions worldwide.

FAQ

What is the difference between OTR and oxygen permeability coefficient?

OTR describes the transmission behavior of a specific specimen of a given thickness under defined conditions, expressed as cm³/(m²·24h). The permeability coefficient is a thickness-normalized material property that allows comparison between materials of different gauges. Both values must always be reported together with the test temperature and relative humidity.

Why do EVOH and nylon films show much higher OTR at high humidity?

EVOH and polyamide are hydrophilic barrier polymers. Absorbed water molecules plasticize the polymer matrix, increasing chain mobility and free volume, which dramatically reduces barrier performance. EVOH can lose more than half of its barrier capacity at 100% RH compared with dry conditions. For this reason, EVOH barrier layers are typically sandwiched between polyolefin layers in real packaging structures, and OTR testing must include the relevant humidity condition.

How long does a typical oxygen permeability test take?

Test duration depends on the barrier level. High-transmission films (OTR above 100 cm³/(m²·24h)) may reach steady state within 2–6 hours. High-barrier materials (OTR below 1 cm³/(m²·24h)) may require 24–72 hours or longer to equilibrate, especially for thick or multilayer laminates with long lag times.

Can finished packages be tested instead of flat sheeting?

Yes. Package-level oxygen ingress testing is covered by ASTM F1307 (coulometric determination of oxygen transmission of rigid packages) and related whole-package methods. However, flat-sheet OTR testing per ASTM D3985 remains the standard approach for material characterization and specification.

What sample quality issues can invalidate test results?

Pinholes, creases, edge leaks, poor sealing, contamination, and non-representative thickness variation are the most common causes of erroneous OTR values. Leak verification, careful specimen cutting, and clean handling procedures are essential for reproducible results.

How often should the instrument be calibrated?

Best practice is to verify instrument calibration with a certified reference film before and after critical test campaigns, and to perform full calibration on a scheduled basis per the manufacturer's recommendations and laboratory quality system requirements (for example, ISO/IEC 17025-aligned procedures).